Chapter 6 – Maintenance
44
Use qualitatively high-grade feed water
Observe the following when using →feed water for the steam
sterilization:
■ Use only →demineralised or distilled water according to →DIN EN
13060, Appendix C.
In case of non-observance, stains or damage on the instruments
can arise and the functional readiness of the steam sterilizer can be
adversely affected.
PLEASE NOTE
→Feed water which you can inexpensively produce with the ion exchanger
MELA
dem
40 or with the revers osmosis process with MELAdem
47,
satisfies the requirements on the feed water.
If the door gasket shrinks, is very wavy or shows cracks in the sealing lip,
it is necessary to exchange the door
The consequence can be leaks which can lead to the leakage of steam.
Safe and successful sterilization may be impacted.
inserting the door gasket it is mandatory to observe the instructions
gasket spare part with the article no. 12550.
Remove the worn door gasket from the groove.
Insert the new door gasket in accordance with the instructions for
fitting the door gasket.
■ Maintenance should only be performed by trained customer services
technicians, or stockist technicians. Consult your specialist dealer or
the nearest MELAG customer services point.
■ Maintain the specified servicing intervals.
Malfunctions at the steam sterilizer can occur on continuation of
operation despite the servicing message.
For retention of value and
functional reliability
Regular maintenance is indispensable for the retention of value and the
reliable practice operation of the
-relevant components and electrical units are
checked during maintenance and replaced where necessary. Maintenance
is carried out according to the maintenance instructions relevant for this
steam sterilizer
Regularly allow for maintenance every 3000 program cycles or every two
years.