Chapter 3 – Initial Commissioning
20
Purified water from the reverse
osmosis unit MELAdem
Drain via waste water container
Fig. 11: Feed water supply from the reverse osmosis unit MELAdem
47
Safeguard combination, consisting of return flow inhibitor and pipe aerator according to EN 1717
Y piece 3/4” (Art. No. 37300)
Cooling water inflow fitting
Leak monitor with shut-off valve and probe (Art.-No. 01056 – optional), recommended with fixed water
connection
Connection for water level monitoring in the waste water container
Water tap (available on-site)
T-piece with hose (Art.-No. 38605)
Feed water filter MELAdem
Cooling water outlet fitting
Reverse-osmosis unit MELAdem
47
Double hose nozzle with anti-flooding flap (Art. No. 37400)
Double-chamber trap (Art. No. 26635)
Water tap DN40 (available on-site)