7 Various functions
780/781 pH/Ion Meter, Manual
93
7.1.3 Addition/Subtraction data (781 pH/Ion Meter only)
ADD.DATA
├ ion type
├ meas.type
├ electr.id
├ variance
├ slope
├ E(0)
├ c Ion
├ date
├ method
├ temperature
├ temp.sensor
├ V total
├ conc.std
├ initial voltage
├ factor
├ smpl.size
├ add1
├ add2
:
├ addn
curve
The data of the last standard or sample addition and subtraction to
have been carried out can be called up with <
> provided
that the measurement type has been set to '
std.add' or 'smpl.add' un-
der
PARAMETER Conc/measuring type. The data are still present
even if the last result has been replaced by '
------' in the measured
value display because the addition run was terminated prematurely.
However, the production of a report of the last result is not possible
then.
The term 'addition' is used below to include both standard and sample
addition and subtraction.
ion type:
The analyte defined for the addition under PARAMETER Conc/ion pa-
rameters
is shown here.
meas.type:
Whether the shown data come from an addition or subtraction run is
shown here.
electr.id
If an electrode Id has been entered or selected before the calibration
under
PARAMETER Conc/measuring parameters/electr.id then it
will be shown here.
Later alterations of the parameters
ion and electr.id have no effects
on the display of the addition data.