7.1 Calibration and addition/subtraction data
92
780/781 pH/Ion Meter, Manual
Although the electrode Id is retained, it cannot be selected under
PARAMETER Conc/measuring parameters/electr.id until it
has been calibrated again.
st1, st2...stn
For each calibration standard the concentration and the measured po-
tential are shown.
A standard which has been deleted by
cal.table: delete n is
shown here by '
deleted'.
curve < –> >
If at least 2 calibration standards have been used then the calibration
curve will be shown as a graph here. The graphics display is accessed
with <
ļ‚’
>:
The display is exited with <
QUIT
>.