7 Various functions
780/781 pH/Ion Meter, Manual
101
7.2.4 Calibration report
In the pH and Conc(direct) modes a calibration report can be produced
after a calibration has been carried out. Under
PARAMETER/calibra-
tion parameters/report
you can define whether this report is to be
produced automatically after the measurement. You can also choose
between a short version containing all the important data and a long
version which additionally contains a plot of the calibration function.
781 pH/Ion Meter 01104 5.781.0020
buffer 1 7.019 0.1 -0.015
buffer 2 3.991 176.7 0.006
buffer 3 9.037 -115.4 0.009
*) dpH: Nominal pH value of the buffer – pH value which, at the
mesured voltage, results from the regression analysis