7 Various functions
780/781 pH/Ion Meter, Manual
101
7.2.4 Calibration report
In the pH and Conc(direct) modes a calibration report can be produced
after a calibration has been carried out. Under
PARAMETER/calibra-
tion parameters/report
you can define whether this report is to be
produced automatically after the measurement. You can also choose
between a short version containing all the important data and a long
version which additionally contains a plot of the calibration function.
'fp
781 pH/Ion Meter 01104 5.781.0020
date 2004-06-02 17:10:16
user C. Weber
pH pH-1 run number 9
electr.id pH electr.01
pH calibration
temperature 20.1 °C
temp.sensor Pt1000
c.dat. 2004-06-02 17:06
buffer type Metrohm
pH U/mV dpH*)
buffer 1 7.019 0.1 -0.015
buffer 2 3.991 176.7 0.006
buffer 3 9.037 -115.4 0.009
variance 1.192
slope 99.44 %
pH(0) 7.035
==============
*) dpH: Nominal pH value of the buffer pH value which, at the
mesured voltage, results from the regression analysis
REPORT
9
+
CAL. DATA
7