EasyManua.ls Logo

Metrohm 780 - Parameter Report

Metrohm 780
177 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.2 Reports
104
780/781 pH/Ion Meter, Manual
The settings shown for the instrument configuration correspond to the
standard settings in the normal operating condition of the pH/Ion Meter
after the configuration memory has been initialized (see Section 8.5) fol-
lowed by conversion to English as the dialog language.
7.2.7 Parameter report
in order to document the parameter settings for the current method you
can produce a parameter report:
'pa
781 pH/Ion Meter 01104 5.781.0020
date 2004-06-03 12:08:36
PARAMETER pH
>measuring parameters
electr.id
drift 0.050 /min
method ********
delta measurement: OFF
stirrer: OFF
>calibration parameters
drift 0.5 mV/min
report: OFF
cal.interval OFF h
no.of buffers 2
buffer type: Metrohm
low lim.slope 95.00 %
up lim.slope 103.0 %
low lim.pH(0) 6.40
up lim.pH(0) 8.00
offset Uoff state: OFF
>limits pH
state: OFF
>limits T
state: OFF
>plot parameters
left mar.: 0.0
right mar.: 14.0
left marg.T 20 °C
right marg.T 30 °C
>preselections
req.ident: OFF
>electrode test
electrode type: standard
report: OFF
--------------
The standard parameters following a memory initialization in the current
mode are shown; in this case the pH mode with a connected tempera-
ture sensor (see Section 8.5). Such parameters can be stored as a
method (see Section 6.1).
PARAM
REPORT
9
+

Table of Contents

Related product manuals