7 Various functions
780/781 pH/Ion Meter, Manual
99
7.2.2 Report identification
The report identification, known here as Report Id, is an abbreviation
describing the type of report and is primarily important for the remote
control of the pH/Ion Meter via the RS232 interface. Computer pro-
grams such as Metrohm Vesuv
©
3.0 for Windows
TM
can identify the
produced report by this identification and use it for further evaluation.
The following report Ids exist for the 780 pH Meter and 781 pH/Ion Me-
ter:
Report Id Report name Availability
'mp
easuring
oints 780, 781
'co
nfiguration 780, 781
'pa
rameter 780, 781
'um
ser
ethods 780, 781
'mv
easured
alues stored 780, 781
'cr pH
alibration
eport 780, 781
'fp
ull
H calibration 780, 781
'di
agnose 780, 781
'et
lectrode
est 780
'cc
onc.
alibration 781
'fc
ull
onc. calibration 781
'ca
onc.
ddition 781
'fa
ull conc.
ddition 781
7.2.3 Measuring point report
By pressing the
key in the directly measuring meas-
urement types a measuring point report will be produced. The following
examples show versions of this type of report, the various configura-
tions of which are defined under
CONFIG/print meas.value (see
Section
5.2).
A typical version of a measuring point report is the output of a single
measured value. The print criterion '
immediate' or 'Drift' must be se-
lected.
781 pH/Ion Meter 01104 5.781.0020
#1 pH = 8.182 21.2 °C (Pt1000)
MEAS/
PRI NT