6 Methods / Parameters
780/781 pH/Ion Meter, Manual
65
6.3 Temperature measurement (T mode)
6.3.1 Measuring parameters
PARAMETER T
├ measuring parameters
: ├ electr.id
├ drift
├ method
├ delta measurement
stirrer
The measuring parameters describe the settings of direct importance
for the measurement.
electr.id 12 ASCII characters
You can assign any name you want to the connected temperature sen-
sor. This sensor Id is then contained in the report header for measured
values.
drift 0.5...1.0...999.9 °C/min, OFF
A measuring signal which changes as a function of time is regarded as
being sufficiently constant when its drift drops below the value defined
here. This can be used as criterion for the pH/Ion Meter if a displayed
measured value is to be stored or documented by printing a report (see
print criterion and storage criterion in Section 5.2 and Section 5.3).
Please note that for a drift-controlled measured value recording this drift
parameter must not be set to '
OFF', as otherwise no result will be ex-
ported.
If this drift condition is not fulfilled then this is indicated by two constant-
ly blinking triangles in the display.
An active drift control increases the reproducibility of the measurements
as a constant condition is used to define the measured result. A smaller
value for the measuring parameter Drift requires a more stable measur-
ing signal than a larger one. The set measuring parameter is also pro-
portional to the measuring uncertainty of the measured value obtained.
method 8 ASCII characters
All parameter settings can be stored together or loaded as a method
(see Section 6.1). The name of the current method is then given here.
This field is only used for information here and cannot be edited direct-
ly.
delta measurement: ON, OFF
The delta measurement permits the display of the measured value rela-
tive to a freely selectable constant reference value:
reference -999.9...0.0...999.9 °C