7.2 Reports
106
780/781 pH/Ion Meter, Manual
7.2.9 User method report
Method and calibration data occupy memory space in the pH/Ion Me-
ter. The user method report gives an overview of the memory occupan-
cy:
'um
781 pH/Ion Meter 01104 5.781.0020
date 2004-06-03 13:58:49
user memory
>methods
pH pH-Demo 82
Conc F(-1) F-Determ 64
>caldata
pH pH 1.1 130
Conc F(-1) F-Sens 146
remaining bytes 4578
--------------
The mode, method name and the occupied memory is given in bytes
for each of the methods. For calibration data the mode, electrode Id
and memory occupancy are given.
A total of 5000 bytes is available for this data. This means that you can
store a maximum of approx. 50 methods or calibration data. This num-
ber can vary depending on the size of the individual data records.
The size of the measured value memory (see Section 7.3) is independ-
ent of this memory administration.