7 Various functions
780/781 pH/Ion Meter, Manual
111
7.5 Setup
Certain settings of the pH/Ion Meter can be found in the
menu.
This menu cannot be accessed from the normal instrument display, as
alterations to these settings have a great influence on the functions of
the pH/Ion Meter.
The
menu is called up immediately after the instrument has been
switched on by pressing down and holding down the
key.
setup
├ lock
├ input assign
├ graphics
Be very careful when making alterations in this menu! Alterations have
a very large influence on the functions of the instrument. Accidental
alterations can be rejected by carrying out a deliberate memory initial-
ization (see Section 8.5).
7.5.1 Locking
setup
├ lock
: ├ all keys
├ <CONFIG>
├ <PARAM>
├ <CAL>
├ recall method
├ store method
├ delete method
├ <CAL.DATA>
├ <MODE>
<EL.TEST>
You can disable certain keys and functions, e.g. to prevent unwanted
alterations to the instrument settings or method parameters.
You can activate such a disablement by setting the corresponding pa-
rameter to '
ON'. This does not affect access to the Setup menu in any
way.
Please note that configuration settings (i.e. CONFIG/report/id1 and
id2) and parameters (i.e. manual calibration concentrations or stand-
ard additions), which are edited during a measurement or calibration
procedure, are modified that way even if
or
are locked. The advantage is that the requested settings and parame-
ters only have to be confirmed during repeated procedures.
Please note that all disablements will be invalidated if a RAM-Init is car-
ried out (see Section 8.5).