4.2 Setting up the titration cell for coulometry
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
899 Coulometer
4.2 Setting up the titration cell for coulometry
4.2.1 Inserting the titration cell
6.2047.030
6.1464.320
1
2
Figure 3 Inserting the titration cell (coulometry)
Set up the titration cell on the magnetic stirrer as follows:
1
Attach the 6.2047.030 titration vessel holder to the magnetic stirrer.
2
Insert the 6.1464.320 titration cell into the titration vessel holder.
4.2.2 Preparing the titration cell
Filling the adsorber tube
Before setting up the titration cell, the 6.1403.030 adsorber tube has to
be filled with 6.2811.000 molecular sieve. Proceed as follows:

Table of Contents