5.9 Live modifications
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
899 Coulometer
5.9.2 Editing the sample table while a determination is running
You can insert new lines or delete existing ones or edit sample data while
a determination is running.
NOTE
We recommend that the editing dialogs always be closed in order to
ensure that no problems occur during the run and that the current data
is always available for calculation purposes.
Editing the sample table
Proceed as follows to edit the sample table:
1
Displaying the main dialog
Press [BACK].
The main dialog is displayed. The determination continues to run in
the background.
2
Opening the main menu
Select Menu and press [OK].
3
Selecting the sample data
Select the menu item Sample table and press [OK].
Select the desired line.
In the function bar, select Edit and press [OK].
4
Editing the sample data
Edit the sample data and apply with Accept or [BACK].
NOTE
In addition to the sample data, the method can also be modified,
except in cases where the determination is running.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 899 Coulometer and is the answer not in the manual?

Metrohm 899 Coulometer Specifications

General IconGeneral
BrandMetrohm
Model899 Coulometer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals