EasyManuals Logo

Metrohm 899 Coulometer User Manual

Metrohm 899 Coulometer
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
■■■■■■■■■■■■■■■■■■■■■■
5 Operation
899 Coulometer
■■■■■■■■
37
5.5 Control
Menu Control
In the dialog Control, the settings for the execution of a single determi-
nation or of one sample series are defined.
Sample table
If this parameter is activated, the sample data for a sample series can be
entered in a table (see Chapter 5.7, page 40).
Selection on | off
Default value off
Autostart
If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see Number of autostarts).
Selection on | off
Default value off
NOTE
If you are working with a sample changer with oven module (e.g. 885
Compact Oven SC), then the Autostart parameter must be set to off,
because the sample changer starts the titration.
Number of autostarts
This parameter is visible only when Autostart = on.
Number of automatic starts.
Input range 1 - 50
Selection table
Default value table
table
The number of automatic starts corresponds to the number of samples
in the sample table.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 899 Coulometer and is the answer not in the manual?

Metrohm 899 Coulometer Specifications

General IconGeneral
BrandMetrohm
Model899 Coulometer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals