5.4 Methods
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
899 Coulometer
Variable Description
DDC Time for drift correction
MIM Initial measured value, i.e. measured value prior to the
processing of the start conditions
MIT Initial temperature, i.e. temperature prior to the process-
ing of the start conditions
MCM End measured value
MCT End temperature
DD Duration of the entire determination
5.4 Methods
5.4.1 Method templates
The 899 Coulometer contains method templates which are already config-
ured except for a few parameters.
The following method templates can be selected:
KFC Coulometric Karl Fischer titration.
The blank value is not taken into account for the
calculation.
KFC-Blank Coulometric Karl Fischer titration minus the
blank value.
Blank Coulometric blank value determination.
The method templates differ only in their calculation formulas.
5.4.2 Loading a method template
Proceed as follows to load a method template:
1
Opening method templates
■ In the main dialog, select Method and press [OK].
The method table with the stored methods opens:
2
In the function bar, select New and press [OK].
The list with method templates opens: