6.1 Creating a titration method
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
NOTICE
The calculation is carried out continuously, i.e. the calculation is
updated after every determination that is carried out until the
defined number of determinations has been reached.
Under [More options], define parameters, e.g., Save result in
result table.
See the detailed manual for precise information.
Return with the [ ] fixed key to the command sequence of the
method run.
4
Adjusting the report options
In the default settings, Result report and Curve are defined in the
report options.
Adjust these parameters as needed and supplement them with addi-
tional reports.
See the detailed manual for precise information.
NOTICE
If you have not connected a printer, delete the command
REPORT. If this is not done, then an error message will appear at
the start of the determination.
5
Adjusting additional method parameters
If necessary, adjust parameters in the KFC command such as the
pause or the extraction time.
See the detailed manual for precise information.
6
Activating the statistics
In the main dialog, tap on [Edit parameters].

Table of Contents

Other manuals for Metrohm 917 Coulometer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 Coulometer and is the answer not in the manual?

Metrohm 917 Coulometer Specifications

General IconGeneral
BrandMetrohm
Model917 Coulometer
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals