EasyManua.ls Logo

MicroP XRF-2000 Series - System Configuration; Set Limit; SET Measuring Time

MicroP XRF-2000 Series
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XRF-2000 Series Operating Manual
25
4.1 System Configuration
Refer to chapter 3.1.1
4.2 Set Limit
Lower : Lower limit value for Low Mode
Upper : Upper limit value for Up Mode
Mode
Low : Measured data is bigger than 'Lower value' means good
UP : Measured data is lower than 'Upper value‘ means good
LowUp: Measured data is between 'Lower value' and 'Upper Value'
means good.
Auto Set Limit: Set the limit value automatically from measured data.
Use Ng/Go Color: Display the data result as a color, Green means good,
Red means no good.
Open/Save: Open or Save the limit value file.
Set: Apply limit value.
Cancel: Cancel limit value.
4.3 Set Measuring Time
Sets the measuring time (the sample is exposed to X-rays). The figure below
defines Clock Time vs. Measuring Time (also called "Live-time") and Dead Time.
Measuring Time Dead Time
Clock Time
The Clock Time is corrected for the Dead Time, always meaning: