XRF-2000 Series Operating Manual
36
4.5 Spectrum window
4.5.1 Qualitative Analysis
The objective of qualitative analysis is to identify the elemental component of
unknown substances. The first step in qualitative analysis is spectra acquisition.
Following spectra acquisition, the spectra must be manipulated and studied to
best determine additional spectra acquisition and/or the qualitative answer
desired. This covers guidelines for qualitative analysis including various spectral
manipulations.
Setting acquisition parameters
(High Voltage and Tube Current setting in CAL File)
Since the purpose of qualitative analysis is to identify the components in a
sample, acquisition parameters should be selected to optimize the
identification CAL File. This is done by maximizing the spectral range
covered by the excitation-detection system as well as obtaining sufficient
sensitivity to identify even trace amounts of the elements present in sample.
This is accomplished by optimizing the X-Ray levels. Total CPS on Spectrum
window is a good indication of this optimized level. Voltage (KV) and current
(uA) settings, combined, should allow for Dead Time 30% in PIN Diode
System, 5% in Proportional Counter system.
Guidelines to excitation parameter efficiencies
Place cursor on Main window and click the Main
window then press Alt+Ctrl+Shif simultaneously and
hold these keys and type “pioneer” and Press Enter
key. Then service menu will appear in spectrum
window. Click Service menu on the Spectrum
window.
On the service menu try to get a good spectrum
regarding application by changing all the parameters
such as HV, TC, Primary Filter and collimator etc.
Clicking adjusting TC to Dead time button will
optimize tube current for 30% of dead time(PIN
Diode) , 5%(Proportional Counter).
For more specific excitation conditions, the use of
filters to modify the exciting radiation is
recommended. Filters modify the exciting radiation
and may selectively enhance the excitation of certain
analytes within the sample.