XRF-2000 Series Operating Manual
63
Select Cal. Mode which is just created. Select biggest collimator and set
measuring time as 100 sec. High Voltage must be adjusted regarding measuring
element.
Ex) ZnNi = 35KV, SnPb = 47KV
New calibration will use pre-defined cal mode. Click Start Cal. button, and enter
number of solution standard.