XRF-2000 Series Operating Manual
65
Load reference standard and click start.
Example of Calibration
: Enter or modify the composition value.
Prepare Plating water standards as follows (Example). Plating water standards
must be measured by AAS to verify its concentration.
SnPb #1 = Sn 6g/L+Pb 1g/L
SnPb #2 = Sn 13g/L+Pb 2g/L