EasyManuals Logo

MicroP XRF-2000 Series User Manual

MicroP XRF-2000 Series
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
XRF-2000 Series Operating Manual
65
Load reference standard and click start.
Example of Calibration
: Enter or modify the composition value.
Prepare Plating water standards as follows (Example). Plating water standards
must be measured by AAS to verify its concentration.
SnPb #1 = Sn 6g/L+Pb 1g/L
SnPb #2 = Sn 13g/L+Pb 2g/L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MicroP XRF-2000 Series and is the answer not in the manual?

MicroP XRF-2000 Series Specifications

General IconGeneral
BrandMicroP
ModelXRF-2000 Series
CategoryMeasuring Instruments
LanguageEnglish