Patient Setup Operation
3 - 10 0070-10-0699-02 Accutorr V Operating Instructions
8. Press (16) or (19) to highlight an initial cuff pressure to change.
9. Once the initial cuff pressure is highlighted, press (18) to select it.
10. Press (16) or (19) to change the initial cuff pressure value.
11. Once the desired pressure is displayed, press (18) to set it.
12. Repeat steps 8 through 11 as needed.
13. Once the initial cuff pressure values are set, press (16) or (19)until OK is
selected.
14. Once OK is selected, press (18) to accept the new cuff pressure values and exit
the INITIAL PRESSURE dialog to the NIBP TOOLS dialog.
NOTE: Select CANCEL and then press (18) to cancel the
operation and exit the INITIAL PRESSURE dialog without
changing the initial pressure values.