5-40 Image Optimization
z Section B: it is the horizontal section in fetal face up posture, as shown in the
figure B above.
z Section C: it is the coronal section in fetal face up posture, as shown in the figure
C above.
Tips: the upper part of the 3D image in the D window is corresponding to the
orientation mark on the probe, if the fetal posture is head down (orientating the
mother’s feet), and the orientation mark is orientating the mother’s head, then the
fetus posture is head down in the 3D image, you can make the fetus head up by
rotating the 3D image by clicking [Quick Rot.] to be “180°” in the soft menu.
Wire cage
When you view a 3D/4D image on the display monitor, it’s sometimes difficult to
recognize the orientation. To help, the system displays a three-dimensional drawing
to illustrate the orientation. Of which, the blue plane presents the image acquisition
where started, while the red plane presents the image acquisition where ended,
besides, a yellow plane in the wire cage presents the position of the sectional plane.
See the graphic below:
Wire Cage
CAUTION:
The ultrasound images are provided for reference only, not
for confirming a diagnosis. Please use caution to avoid
misdiagnosis.
5.10.2 Note Before Use
5.10.2.1 3D/4D Image Quality Conditions
NOTE: In accordance with the ALARA (As Low As Reasonably Achievable) principle,
try to shorten the sweeping time after a good 3D image is obtained.
The quality of images rendered in 3D/4D mode is closely related to the fetal condition,
angle of a B tangent plane and scanning technique (only for Smart 3D). The following
description uses fetal face imaging as an example. Imaging of other parts is the same.
Fetal Condition
(1) Gestational age
Fetuses of 24~30 weeks old are the most suitable for 3D imaging.
(2) Fetal body posture
Recommended: cephalic face up (Figure a) or face to the side (Figure b);
NOT recommended: cephalic face down (Figure c).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-40 and is the answer not in the manual?

Mindray DC-40 Specifications

General IconGeneral
Touch ScreenYes
Hard Drive500 GB
3D/4D ImagingYes
Storage Temperature-20°C to +60°C
Operating Temperature+10°C to +40°C
Atmospheric Pressure700 hPa to 1060 hPa
Imaging ModesB, M, Color Doppler, Power Doppler
ProbesConvex, Linear, Phased Array
ConnectivityDICOM, USB, Ethernet
Power Supply100-240V AC, 50/60Hz
Display15-inch color LCD