EMC Guidance and Manufacturer’s Declaration 15-1
15 EMC Guidance and
Manufacturers Declaration
The system complies with the EMC standard IEC 60601-1-2: 2007.
WARNING:
1. The use of unapproved accessories may diminish system
performance.
2. Use of components, accessories, probes, and cables other
than those specified may result in increased emission or
decreased immunity of system.
3. The system or its components should not be used adjacent to
or stacked with other equipment. If adjacent or stacked use is
necessary, the system or its components should be observed
to verify normal operation in the configuration in which it will
be used.
4. Operation of system, in the case that the patient physiological
signal is lower than the minimum amplitude or value specified
in the product specifications, results may be inaccurate
(results can be obtained when the HR is in the range of 30-250
bmp or when the QRS wave amplitude is between 0.5-5 mV.)
NOTE:
1.
The system needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided below.
2.
Other devices may interfere with this system even though they meet the
requirements of CISPR in the corresponding standards.
3.
Preventing conducted RF immunity. Due to technological limitations, the
conducted RF immunity level are limited to 3 Vrms level, conducted RF
interference above 3Vrms may cause wrong diagnosis and measurements. We
suggest that you position system further from sources of conducted RF noise.
4.
Portable and mobile RF communications equipment can affects system. See
tables 1, 2, 3, and 4 below.
If the system is operated within the electromagnetic environment listed in Table 2 and Table 3, the
system will remain safe and will provide the following basic performances:
z Imaging;
z Doppler acoustic spectral displaying;
z Taking measurements;
z Patient information;
z Date/time information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-40 and is the answer not in the manual?

Mindray DC-40 Specifications

General IconGeneral
Touch ScreenYes
Hard Drive500 GB
3D/4D ImagingYes
Storage Temperature-20°C to +60°C
Operating Temperature+10°C to +40°C
Atmospheric Pressure700 hPa to 1060 hPa
Imaging ModesB, M, Color Doppler, Power Doppler
ProbesConvex, Linear, Phased Array
ConnectivityDICOM, USB, Ethernet
Power Supply100-240V AC, 50/60Hz
Display15-inch color LCD

Related product manuals