EasyManuals Logo

Mindray DC-40 User Manual

Mindray DC-40
303 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
8-4 Measurement
Abdomen measurements - Used for measurements of abdominal organs (liver, gall
bladder, pancreas and kidney, etc.) and large abdominal vessels.
OB measurements - Used for measurements of fetal growth indexes (including EFW) as
well as GA and EDD calculations. The fetus can be evaluated through growth graph
analysis and fetal biophysical profile.
Cardiac measurements – Used for left ventricle function measurements and
measurements of main artery and vein parameters, etc.
Gynecology measurements - Used for the uterus, ovary and follicles, etc.
Small Part measurements – Used for small parts such as thyroid.
Urology measurements - Used for prostate, seminal vesicle, renal, adrenal, micturated
and testicle volume.
Orthopedics measurements – Used for hip joint measurement.
Vascular measurements – Used for carotid, cerebral, upper and lower extremities vessels,
etc.
EM measurements – Used for EM related full functional measurement.
8.4 Measurement Accuracy
Table 1 Error of 2D Images
Parameter Value Range Error
Distance Full screen Within ±2%
Area Full screen Within ±6%
Circ Full screen Within ±7%
Angle Full screen Within ±3%.
Volume Full screen Within ±7 %
Table 2 iScape Measurements
Parameter Probe Type Error
Distance
Full screen Within ±5 % (L14-6NE, L12-3E, P4-2, D7-
2E, 3C5A, 7L4A probe)
Full screen Within ±10 % (V11-3, CB10-4E probe)
Table 3 Time/Motion Measurements
Parameter Value Range Error
Distance Full screen Within ±3 %
Time Timeline Display Within ±2%
Heart rate Timeline Display Within ±3%
Velocity (PW
mode)
10-200cm/s: (for
L14-6NE, L12-3E,
CB10-4E, 7L4A)
10-300cm/s: (for
P4-2, D7-2E, V11-3,
3C5A)
When angle60º, 5%.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-40 and is the answer not in the manual?

Mindray DC-40 Specifications

General IconGeneral
Touch ScreenYes
Hard Drive500 GB
3D/4D ImagingYes
Storage Temperature-20°C to +60°C
Operating Temperature+10°C to +40°C
Atmospheric Pressure700 hPa to 1060 hPa
Imaging ModesB, M, Color Doppler, Power Doppler
ProbesConvex, Linear, Phased Array
ConnectivityDICOM, USB, Ethernet
Power Supply100-240V AC, 50/60Hz
Display15-inch color LCD

Related product manuals