Measurement 8-5
Parameter Value Range Error
Velocity (CW
mode)
10-300cm/s (for P4-
2)
When angle60º, 5%. (not including pencil
probe)
Table 4 Doppler Velocity Measurements
Parameter Value Range
Velocity (PW mode)
Min.0.10cm/s, Max.462cm/s
Velocity (CW mode)
Min.0.06cm/s, Max.3080m/s
Velocity (Color mode)
Min.1cm/s, Max..239.5cm/s
NOTE: Within the selected field range, the measurement accuracy is ensured within
the range mentioned above. The accuracy specifications are performance in
the worst conditions, or based on the real test for the system, regardless of
acoustic speed error.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-40 and is the answer not in the manual?

Mindray DC-40 Specifications

General IconGeneral
Touch ScreenYes
Hard Drive500 GB
3D/4D ImagingYes
Storage Temperature-20°C to +60°C
Operating Temperature+10°C to +40°C
Atmospheric Pressure700 hPa to 1060 hPa
Imaging ModesB, M, Color Doppler, Power Doppler
ProbesConvex, Linear, Phased Array
ConnectivityDICOM, USB, Ethernet
Power Supply100-240V AC, 50/60Hz
Display15-inch color LCD