Guidance and Manufacturer’s Declaration 18-3 
TABLE 2 
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY 
The system is intended for use in the electromagnetic environment specified below. The 
customer or the user of system should assure that it is used in such an environment. 
IMMUNITY TEST
IEC 60601 
TEST LEVEL 
COMPLIANCE 
LEVEL 
ELECTROMAGNETIC 
ENVIRONMENT-GUIDANCE
Electrostatic 
Discharge(ESD) 
IEC 61000-4-2 
±6 kV contact; 
±8 kV air 
±6 kV contact; 
±8 kV air 
Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the 
relative humidity should be at 
least 30%. 
Electrical fast 
Transient / burst 
IEC 61000-4-4 
±2 kV for power 
supply lines; 
±1 kV for 
input/output 
lines 
±2 kV for power 
supply lines; 
±1 kV for 
input/output lines 
Mains power quality should 
be that of a typical 
commercial or hospital 
environment. 
Surge 
IEC 61000-4-5 
±1 kV line(s) to 
line(s); 
±2 kV line(s) to 
earth 
±1 kV line(s) to 
line(s); 
±2 kV line(s) to 
earth 
Mains power quality should 
be that of a typical 
commercial or hospital 
environment. 
Voltage dips, 
Short interruptions 
and voltage 
variation on power 
supply input 
voltage   
IEC 61000-4-11 
<5% U
T 
(>95% 
dip in U
T 
) for 
0.5 cycle 
 
40% U
T 
(60% 
dip in U
T 
) for 5 
cycle 
 
70% U
T 
(30% 
dip in U
T 
) for 25 
cycle 
 
<5% U
T 
(>95% 
dip in U
T 
) for 5 
sec 
<5% U
T 
(>95% 
dip in U
T 
) for 0.5 
cycle 
 
40% U
T 
(60% dip 
in U
T 
) for 5 cycle 
 
70% U
T 
(30% dip 
in U
T 
) for 25 cycle
 
<5% U
T 
(>95% 
dip in U
T 
) for 5 
sec 
Mains power quality should 
be that of a typical 
commercial or hospital 
environment. If you require 
continued operation during 
power mains interruptions, it 
is recommended that our 
product be powered from an 
uninterruptible power supply 
or a battery.   
Power frequency 
(50/60 HZ) 
magnetic field 
IEC 61000-4-8 
3 A/m  3 A/m  Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 
environment. 
NOTE: U
T
 is the A.C. mains voltage prior to application of the test level.