18-4 Guidance and Manufacturer’s Declaration 
TABLE 3 
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY 
The system is intended for use in the electromagnetic environment specified below. The 
customer or the user of system should assure that it is used in such an environment. 
IMMUNITY 
TEST 
IEC 60601 
TEST LEVEL
COMPLIANCE 
LEVEL 
ELECTROMAGNETIC 
ENVIRONMENT-GUIDANCE 
Conduced RF 
IEC 61000-4-6 
3 Vrms 
150 kHz - 80 
MHz 
1 Vrms 
Portable and mobile RF 
communications equipment should 
be used no closer to any part of 
system, including cables, than the 
recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. Recommended 
separation distance 
d = 3.5 x P  
d = 1.2 x P   80 MHz to 800 MHz 
d = 2.3 x P   800 MHz to 2.5GHz 
Where, P is the maximum output 
power rating of the transmitter in 
watts (W) according to the 
transmitter manufacturer and d is 
the recommended separation 
distance in meters (m). 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey, should 
be less than the compliance level in 
each frequency range.   
Interference may occur in the 
vicinity of equipment marked with 
the following symbol: 
 
Radiated RF 
IEC 61000-4-3
3 V/m 
80MHz - 
2.5GHz 
3 V/m 
Note 1  At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2  These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 
 
Field strengths from fixed transmitters, such as base stations for radio (cellular 
/cordless) telephones and land mobile radios, amateur radio, AM and FM radio 
broadcast and TV broadcast cannot be predicted theoretically with accuracy. 
To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength 
in the location in which system is used exceeds the applicable RF compliance 
level above, system should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as 
reorienting or relocating system. 
 
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 
1V/m.