Guidance and Manufacturer’s Declaration 18-5
TABLE 4
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE
RF COMMUNICATION DEVICE AND SYSTEM
The system is intended for use in an electromagnetic environment in which radiated RF
disturbance are controlled. The customer or the user of system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communication equipment (transmitters) and system as recommended below,
according to the maximum output power of the communication equipment.
Rated Maximum
Output power of
Transmitter
(W)
Separation Distance According to Frequency of Transmitter
(m)
150kHz -80MHz
Pd






=
1
5.3
80MHz-800MHz
Pd






=
3
5.3
800MHz-2.5GHz
Pd






=
3
7
0.01
0.35 0.12 0.23
0.1
1.11 0.37 0.74
1
3.50 1.17 2.34
10
11.07 3.69 7.38
100
35.00 11.67 23.34
For transmitters at a maximum output power not listed above, the recommended
separation distanced in meters (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
If system image distortion occurs, it may be necessary to position system further from
sources of conducted RF noise or to install external power source filter to minimize RF
noise to an acceptable level.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

Table of Contents

Other manuals for Mindray M7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M7 and is the answer not in the manual?

Mindray M7 Specifications

General IconGeneral
TypePortable Ultrasound System
Display15-inch LCD
Imaging Modes2D, M-mode, Color Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Cardiac, Vascular, OB/GYN, Small Parts, Musculoskeletal, Urology, Pediatric
BatteryRechargeable Lithium-ion
ConnectivityUSB, Ethernet, DICOM
Dimensions35 cm x 36 cm x 9 cm
Transducer Ports2 active transducer ports