EasyManua.ls Logo

Minebea Intec PR 5310 - Page 128

Minebea Intec PR 5310
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Operating | 6.2 - Configuration and adjustment via the keypad | 128
GroSS appears on the display.
6. Press / to select the corresponding mode, see parameter table.
7. Press "OK".
Ao 051 appears on the display.
8. Press "OK".
AnA.rnG appears on the display.
9. Press "OK".
0-20MA appears on the display.
10. Press / to select the corresponding range, see parameter table.
11. Press "OK".
Ao 052 appears on the display.
12. Press "OK".
out.Err appears on the display.
13. Press "OK".
4 MA appears on the display.
14. Press / to select the corresponding output value, see parameter
table.
15. Press "OK".
Ao 053 appears on the display.
16. Press "OK".
out.< 0 appears on the display.
17. Press "OK".
4 MA appears on the display.
18. Press / to select the corresponding output value, see parameter
table.
19. Press "OK".
Ao 054 appears on the display.
20. Press "OK".
out.>20 appears on the display.