4. Device installation | 4.3 - Hardware construction | 45
4.3.3.6 Connecting load cells of type series PR 6221
See installation manuals of PR 6221 and PR 6021/08, ../18, ../68S.
4.3.3.7 Testing the measuring circuit
A simple test with the load cells connected can be carried out with a
multimeter.
Note: In the case of an external load cell supply voltage or use of an
isolating unit, the internal load cell supply is not relevant.
Measuring voltage
0…12mV = @ LC with 1.0mV/V
0…24mV = @ LC with 2.0mV/V