29
Procedures
7
Press the “ENTER” key.
The load display will become “0.0000”, and the
10
0
digit will light on and off.
When the calibration has been already
completed, the output value of load cell
registered at that time will be displayed.
Set the output value of load cell with the initial
load application by using “A/Z A ”, “A/Z OFF Y”
or “CHECK” keys.
Though the number of digits has not
prepared in the “Inspection data” for load
cell so many as shown in the right figure,
extra digits are necessary for the
compensation with the standard point at
internal of the instrument.
In case of actual setting, insert “0” into the
extra digits.
As for the value for extra digit, when tare
compensation and fine adjustment on load
are applied, it will be written as a
compensated value automatically.
Setting range for the output of load cell is
from −0.100 0 mV/V to 2.400 0 mV/V with
initial load applied.
“A/Z A ” key :Set value carry key
“A/Z OFF Y” key :Set value Increment key
“CHECK” key :Set value Initialization key
By pressing the key continuously,
increment can be provided continuously.
8
Press the “ENTER” key.
The load display will show “S MV”.