30
Procedures
Press the “ENTER key.
The load display will become “3.0000”, and the
digit of 10
0
will light on and off.
When calibration has been already completed,
the output value of load cell registered at that
time will be shown.
Set the output value for load cell corresponding
to the maximum display value set in step 5 with
the A/Z A”, A/Z OFF Y or “CHECK” key
9
The setting value to be set here should be
at 0.2 mV/V or more than the set value in
the step 7.
Though the number of digits has not
prepared in the “Inspection data” for load
cell as many as the digits in the right
figure, extra digits are necessary for the
compensation for the internal standard
point of the instrument.
In case of actual setting, insert “0” into
the extra digits.
As for the value of extra digit, when tare
compensation and fine adjustment on load
are applied, it will be written as a
compensated value automatically.
A/Z A key Set value carry key
A/Z OFF Y key Set value Increment key
“CHECK” key Set value Initialization key
By pressing the key continuously,
increment can be provided continuously.
10
Press the “ENTER key.
The load display will show “TEND”.
11
Press the “ENTER key.
Quitting from the ACAL mode, the present load
value will show on the load display.