35
Procedures
Press the “ENTER” key.
The load display will become “−−−−−” with
flashing display on and off, then zero adjustment
can be started.
9
Warning : At the same time, be careful
not to apply load vibration due to
vibration and so on.
When load vibration is applied, it may
have the possibilities that zero point
becomes unstable, and precise reading of
zero will not be obtained.
After completing above, the indication on load
display will become “SPAN”.
However, when the initial load is not entered
within the range from −0.1 mV/V to 2.4 mV/V,
the error code shown on the right figure will be
displayed for about 2 s, and the load display will
show “ZERO”, then will return to step 8.
Error code
TER−L :Zero point − over
TER−H :Zero point + over
Error code
10
Apply th
e
same weight on th
e
load cell as se
t
in
the step 7.