79
D F−72 Actual load value (For reference)
※The actual load value set when calibration (LCAL) is applied can be referred
to.
※When the calibration except LCAL is made, this value will not change.
※In this function, change of setting can’t be made.
D F−73 Zero calibration value (For reference)
※The input voltage value which is read as initial load value at the time of
performing the calibration can be referred to.
※In this function, change of setting can’t be made.
※Since there is no decimal point on the display, think the value as if it had
the decimal point as follows:
(ex.)When the display shows “00000”, it should be treated as “0.0000”.
D F−74 Span calibration value (For reference)
※The input voltage value at the time of maximum display can be referred to.
※In this function, change of setting can’t be made.
※Since the display doesn’t have the decimal point, think the value as if it had
the decimal point as follows:
(ex.)When the display shows “30000”, it should be treated as “3.0000”.
D F−97 Setting data width of stabilized filter
Initial value=00020 00000=Stabilized filter OFF
Setting range:00000〜00999
Unit:1D
Data width of 999D with the setting of “00999”.
※Effective only when setting is made between 00001〜00002 with F−09, and
between 00001〜00099 with F−10 as well.
D F−98 ZERO clear
Zero compensated data by zero set function, compensated data by Tare weight
cancellation (A/Z) function and compensated data by zero tracking can be
cancelled. When the “ENTER” key is pressed with “F−98” displayed, “ZCLR?”
display can be obtained. (At the same time, the display lights on and off.)
At this point, press the “FUNC.” key when suspending the ZERO clear is
desired. Measurement mode can be returned and Memory clear will not be
executed. When the “ENTER” key is pressed while “ZCLR?” display lights on
and off, “F−98” display can be returned. Now, ZERO clear has completed.
D F−99 Memory clear
Setting from F−01 to F−97 recorded at EEPROM will return to the default
value. When the “ENTER” key is pressed with the display of F−
99, then
“CLR?” display can be obtained. (At the same time, the display light on and
off.) At this point, press the “FUNC.” key when suspending memory clear is
desired. Measurement mode can be returned and Memory clear will not be
executed. When the “ENTER” key is pressed during “CLR?” load display
lights on and off, and after about 2 s, it will become “FUNC” display and
operation of Memory clear has completed.