78
D F−21 Selection of output logic of BCD data
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−22 Selection of output logic of BCD polarity (POL)
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−23 Selection of output logic for BCD flag(ERROR、OVR)
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−43 Setting CHECK value
Initial value=00003 Setting range 00000〜00015
Unit:0.1 mV/V
Approx. 1.5 mV/V of CHECK value with the setting
of “00015”.
D F−50 Selection of the target of HOLD
Initial value=00003 00000=Ineffective for HOLD.
00001=HOLD only for display
00002=HOLD only for BCD output
00003=HOLD for both of display and BCD output
D F−59 Selection of digital Tare weight cancellation
Initial value=00000 00000=Ineffective
00001=Effective
※Selects whether digital Tare weight cancellation is to be effective or
ineffective.
D F−60 Setting digital Tare weight cancellation value
Initial value=00000 Setting range −99990 to 99990
※When digital Tare weight cancellation is to be “Effective” with F−
59, digital
Tare cancellation for the set portion by this setting can be executed.
※The following calculation should be made for digital Tare weight
cancellation against “Gross weight”.
(Gross weight)−(Digital Tare weight cancellation value)
=(Display when digital Tare cancellation is effective.)
D F−62 Setting the target of zero tracking
Initial value=00000 00000=Gross weight
00001=Net weight
D F−70 Increment value (For reference)
※The increment value set when the calibration is applied can be referred to.
※In this function, change of setting can’t be made.
D F−71 Maximum display value (For reference)
※The maximum display value set when the calibration is applied can be
referred to.
※In this function, change of setting can’t be made.