78
D F21 Selection of output logic of BCD data
Initial value00000 00000Negative logic
00001Positive logic
D F22 Selection of output logic of BCD polarity (POL)
Initial value00000 00000Negative logic
00001Positive logic
D F23 Selection of output logic for BCD flag(ERROROVR)
Initial value00000 00000Negative logic
00001Positive logic
D F43 Setting CHECK value
Initial value00003 Setting range 0000000015
Unit0.1 mV/V
Approx. 1.5 mV/V of CHECK value with the setting
of “00015”.
D F50 Selection of the target of HOLD
Initial value00003 00000Ineffective for HOLD.
00001HOLD only for display
00002HOLD only for BCD output
00003HOLD for both of display and BCD output
D F59 Selection of digital Tare weight cancellation
Initial value00000 00000Ineffective
00001Effective
Selects whether digital Tare weight cancellation is to be effective or
ineffective.
D F60 Setting digital Tare weight cancellation value
Initial value00000 Setting range 99990 to 99990
When digital Tare weight cancellation is to be “Effective” with F
59, digital
Tare cancellation for the set portion by this setting can be executed.
The following calculation should be made for digital Tare weight
cancellation against “Gross weight”.
Gross weight)−(Digital Tare weight cancellation value
=(Display when digital Tare cancellation is effective.)
D F62 Setting the target of zero tracking
Initial value00000 00000Gross weight
00001Net weight
D F70 Increment value (For reference)
The increment value set when the calibration is applied can be referred to.
In this function, change of setting can’t be made.
D F71 Maximum display value (For reference)
The maximum display value set when the calibration is applied can be
referred to.
In this function, change of setting can’t be made.