EasyManua.ls Logo

MKS 640A/641A - Figure 11: Exploded View of the W-Seal Components; How to Connect the W-Seal (SEMI 2878.3) Fittings

MKS 640A/641A
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Down Port Controller Mounting Instructions Cha
p
ter Two: Installation
26
How To Connect the W-Seal (SEMI 2878.3) Fittings
Note
The
g
askets used in the W-seal are also referred to as IGS (Inte
g
rated Gas
Systems) gaskets. Replacement gaskets (part number UGF-6.35G) are
available from Fujikin of America, Inc.
1.
Remove the protective packaging from the controller.
Fujikin has a gasket setting tool (part number IGS-Gasket-tool) available to facilitate the
assembly process.
2.
Hold the W-seal gasket by the retainer and remove the protective packaging.
3.
Use filtered nitrogen to blow clean the controller and W-seal gasket.
Caution
Be careful not to damage the gasket. A damaged gasket will not seal
properly.
4.
Install the gasket and retainer into the controller.
The retainer will hold the gasket in place.
M5-0.8 x 30 mm Long
Mounting Screw (1 of 4)
Leak Detection Port
W-Seal Gasket
and Retainer Assembly
Gasket and Retainer
Assembly Exploded
View
Figure 11: Exploded View of the W-Seal Components

Table of Contents