Dansensor® MAP Check 3 Vacuum EN User Guide 35
P/N 320598-G
11/2019
Text colors indicate the following:
Black indicates that the result is within limit or that no limit
has been defined for this gas.
Orange indicates that the gas concentration has exceeded
the set warning value.
The arrow symbol to the right of the value indicates
whether it is an upper or lower warning.
Red indicates that the gas concentration has exceeded
the set alarm value.
The arrow symbol to the right of the value indicates
whether it is an upper or lower alarm.
Status Shows current status of the device e.g. “Ready” or “Measuring”
Faults disabled This symbol indicates that faults has been disabled in
“Measure setup”.
All faults are detected but external signalling is disabled.
This is used as last resort if you want to be able to run
packaging machine even though an error exists in the device.
Product key Takes you to the “Products” list for selection of product.
See "Selecting a product for measuring" on page 36 for details.
Menu key Takes you to the “Main menu”.
See "Main menu" on page 53 for details.
Measure/Stop key Starts/stops measurement.
Mixer connection (Only appears if External mixer is enabled)
Shows current connection status of the external mixer.
= Connection OK
= No connection
Pressing the icon changes the screen to display the mixer flow,
mix and pressure readings.
Cycle count Shows the number of completed packaging machine cycles
Buffer measurement (”Buffer” versions only)
Displays the result of the latest buffer measurement. Read-out
includes time of measurement, O
2
/CO
2
values and buffer
pressure.
Text colors indicate the following:
Black indicates that the result is within limit or that no limit
has been defined for the gasses and the pressure.
Orange indicates that the result has exceeded the set
warning value.
Red indicates that the result has exceeded the set alarm
value.
The small display is also a key that you can press to start a
buffer measurement manually.