34 User Guide EN Dansensor® MAP Check 3 Vacuum
P/N 320598-G
11/2019
Measuring
"Black Box" models
The “BlackBox” models are as default set to "Autostart". When heating period has finished the
device starts automatically and checks for the “Evacuation” and “Gassing” signals. When
receiving a signal the device starts a cycle. A new signal is required to start the next cycle and
so on.
Display models
When set to "Autostart" the “Display” models remember how the device was setup when it
was turned off. If the device was measuring then it will automatically start to check for the
“Evacuation” and “Gassing” signals when heating period has finished. When receiving a signal
the device starts a cycle. A new signal is required to start the next cycle and so on.
If the device was stopped then it will stop when heating period finishes.
It is recommended is to leave the device in measure mode so that it starts automatically.
Alternatively the measuring can be started and stopped manually using the Measure/Stop
key in the measuring screen - see "The Measuring Screen" below.
The Measuring Screen
CAUTION! The touch screen should only be finger touch operated. Pencils or metal
tools will destroy the touch sensitive film.
The measuring screen offers the following information/functions:
Selected product Currently selected product number/product name.
Remote control Displayed when the device is remote controlled from Modbus
TCP protocol.
In this case the buttons ,,and are disabled.
Time/Date Current time and date
Measuring result Result of the recent measurement.
During heating up the remaining heating time is also
displayed in this area.
Only values selected for display in the product setup are
displayed - see
"Edit product" on page 54 for details.