Test modes in Diagnostics
Test measuring - Cal inlet
This test is used for calibration of the sensors in the system. It is normally used for servicing
purposes.
Remove the blind plug from the input labelled “CAL. GAS IN” and connect a suitable M5
fitting.
Connect a calibration gas to the input.
The gas must be regulated using a needle valve for a flow of 0.5 L/min ± 0.1 L/min.
Apply vacuum to the device and stop the packaging machine.
Enter Diagnostics and go to page 2 and select Test measuring - Cal inlet in the menu.
The device starts a continuous measurement of the connected calibration gas.
It is now possible to close the Diagnostics window and to go to the measuring screen to
watch the gas values.
The measuring screen will say “Test measuring”.
The test will run until you press Stop or until Test measuring - Cal inlet is pressed again on
page 2 in Diagnostics.
NOTE! Always install the blind plug in the “CAL. GAS IN” input when not in use.