Dansensor® MAP Check 3 Vacuum EN User Guide 45
P/N 320598-G
11/2019
Cleaning and Maintenance
General
Performing cleaning and maintenance regularly reduces the chances of equipment failure.
CAUTION! Personnel performing any maintenance or cleaning must familiarize
themselves with the "Safety Instructions" on page 9 before attempting any of these
procedures.
NOTE! For correct mounting/dismounting of fittings please see "Correct mounting/
dismounting of fittings" on page 19.
Spare parts
For a list of spare parts see "Spare parts, consumables and accessories" on page 87.
Cleaning
All of the device’s surfaces should only be cleaned using a mild soap solution and a wrung
cloth.
CAUTION! Never use hard tools or abrasive materials when cleaning any part of the
device.
WARNING! Never use cleaning agents containing chlorinated solvents or acetic or
phosphoric acid. These constitute a health hazard and could damage the
instrument.
Replacing filter in the sample gas probe
We recommend that you inspect the filter regularly and replace it when necessary.
If the filter is blocked the device will indicate an error in the display.
To replace the filter simply unscrew the connection fittings.