Dansensor® MAP Check 3 Vacuum EN User Guide 49
P/N 320598-G
11/2019
Test measuring - Buffer inlet
This test is used for a manual continuous measurement of the buffer tank contents. It is
normally used for servicing purposes or to check the tank after installation.
Apply vacuum to the device and stop the packaging machine.
Enter Diagnostics and go to page 2 and select Test measuring - Buffer inlet in the menu.
The device starts a continuous measurement of the buffer tank gas.
It is now possible to close the Diagnostics window and to go to the measuring screen to
watch the gas values.
The measuring screen will say “Test buffer measuring”.
The test will run until you press Stop or until Test measuring - Buffer inlet is pressed again
on page 2 in Diagnostics.
25: Meat Mix 1000gr.
Cycles
6250
Tes t bu ff er
measuring
O
2
Bal
CO
2
%
%
%
Buffer
09:42
%O
2
%CO
2
bar
21,0
0,0
2,2