84 User Guide EN Dansensor® MAP Check 3 Vacuum
P/N 320598-G
11/2019
Gas connection
Measurement gas inlet Hose Ø5/3 mm, sample hose l=3m
Vacuum connection Hose Ø8/6 mm
Buffer gas inlet Push-in, OD=3 mm
Calibration gas inlet M5 female (normally plugged)
Air inlet/Cal. gas overflow M5 female (normally fitted with filter)
Basic specifications
Warm-up time Operational after 10 min.
(60 minutes to full specifications)
Measuring ranges 0-100% O
2
and optionally 0-100% CO
2
Ambient temperature Operational: 0 to +45 °C, less than 95% RH,
non-condensing
Storage: -10°C to +60°C, less than 95% RH,
non-condensing
Ambient pressure Operational:900 hPa to 1050 hPa
Measurement gas Inert gasses (O
2
, CO
2
, N
2
, Ar, Air), less than 95% RH
Measurement pressure Measure input: 0-1050 mbar
1
Buffer versions - gas inlet: 1.5-10 bar
Vacuum < 100 mbar abs. @ flow of 5 l/min.
1
Vacuum >= 400 mbar
O
2
sensor
Sensor type Zirconia
Measure gas temperature +0 to +35 °C
Measure flow Measure input: Typically 5 l/min. ± 0.5 l/min.
Sensor flow: 75 ml/min. ± 5 ml/min.
Buffer measurement: 75 ml/min. ± 5 ml/min.