Dansensor® MAP Check 3 Vacuum EN User Guide 59
P/N 320598-G
11/2019
Diagnostics menu
Selecting Diagnostics menu from the Main menu will take you to the device’s Diagnostics
pages.
The Diagnostics section consists of 3 pages. Use the arrow keys in the headline area to go to
next or previous page:
The first page shows the internal parameters of the device. These are values such as the
device’s internal temperature, the pressure in the sensors, the sensor’s conditions, serial
numbers for main components, day/hour counters, time to calibration, and software
versions, etc. The parameters can only be read and not changed.
The second page holds a menu from where you can
-open the Diagnostics error log and Diagnostics events log pages.
- start a test measuring on the Cal. Gas or Buffer Gas inlets.
The measure functions are normally used for initial testing after installation on machine.
-start a System leak test.
The third page shows the state of the valves, the Vac and Gas signals, and the Ready,
Warning and Alarm relays.