Dansensor® MAP Check 3 Vacuum EN User Guide 85
P/N 320598-G
11/2019
CO
2
sensor
Sensor type Infrared, NDIR, temperature controlled to +60 °C
Measure gas temperature +0 to +35 °C
Measure flow
2
Measure input: Typically 5 l/min. ± 0.5 l/min.
Sensor flow: 75 ml/min. ± 5 ml/min.
Buffer measurement: 75 ml/min. ± 5 ml/min.
2
At ambient pressure of 1013 mbar and vacuum < 100 mbar abs
Accuracy specification (excl. calibration)
Zirconia sensor resolution 0.1% absolute in range above 10%
0.01% absolute in range above 1%
0.001% absolute in range below 1%
Zirconia sensor accuracy <0.01% absolute in range below 0.01%
3
±0.001% absolute in range below 0.1%
±1% relative in range above 0.1%
CO
2
sensor resolution 0.1% absolute
CO
2
sensor accuracy ±0.5% absolute ± 1.5% of reading
NOTE! High concentrations of Argon influences
accuracy of CO
2
readings. The CO
2
value will appear to
be slightly lower than actual value (app. -2-4%).
3
Not calibrated below 1000 ppm
NOTE! Accuracy specifications are valid at the “Specification conditions” (see page
86).