60 User Guide EN Dansensor® MAP Check 3 Vacuum
P/N 320598-G
11/2019
General setup
Selecting General setup from the Main menu will display a menu with available setup
parameters.
“Supervisor” access level is required for this menu, as the parameters in this menu control the
basic functionality of the device.
The General setup menu holds the following items:
Autostart No Device must be started manually using the
“Measure” button on the display
Ye s The device starts automatically when heating
period ends.
Disable faults No All faults are detected and READY relay contacts
will be activated to stop packaging machine logic.
Ye s All faults are detected but external signalling is
disabled.
This is used as last resort if you want to be able to
run packaging machine even though an error
exists in the device.
This setting will be indicated in the measure
screen by a warning symbol and a red text saying
"Faults disabled".
External mixer settings Opens the External mixer settings screen
See "External mixer settings" on page 61 for details.
Buffer measuring setup Opens the Buffer measure setup screen
See "Buffer measure setup" on page 62 for details.
Cycle counter Can be set to any number. Counter is set to 0 from the factory.
Can be set to match the packaging machine cycle counter.
Data log setup Opens the Data log setup screen
See "Data log setup" on page 63 for details.
Network setup Opens the Network setup screen
See "Network setup" on page 68 for details.
User label: This item is a user configurable text that will be combined with
data logging information. Examples of use could be
manufacturing line identification or other similar data.
See "User label:" on page 68 for details.
General setup