86 User Guide EN Dansensor® MAP Check 3 Vacuum
P/N 320598-G
11/2019
Standard calibration specification
Calibration gasses (Zr O
2
) 1000 ppm, 1%, 80% (balance N
2
)
20.9% (Compressed dried atmosphere air)
Calibration gasses (CO
2
) 0%, 25%, 60%, 100% (balance N
2
)
Calibration gas accuracy < 3%
System diffusion < 25 ppm
Sensor gas flow 75 ml/min. ± 5 ml/min. (at 20.9% O
2
/ Bal. N
2
)
Ambient temperature 25 °C ± 5 °C
Ambient relative humidity 35 - 65% RH
Ambient pressure 1013 hPa ± 50 hPa
Total calibration accuracy (RMS) ± (25 ppm + 4% of reading) in the range
1000ppm - 100%
Specification conditions
Ambient temperature +25 °C
Ambient pressure 1013 hPa
Measurement gas temperature +23 °C
Ambient relative humidity 40% RH
Measurement gas relative humidity <10% RH
Sensor gas flow 75 ml/min
Vacuum <25 mbar abs
NOTE! All gas concentrations are specified in volume percent.
Conformity
CE
RoHS
China RoHS Phase 1 compliance
Food regulation 1935/2004