EasyManua.ls Logo

Mocon AMETEK Dansensor MAP Check 3 Vacuum - Page 57

Mocon AMETEK Dansensor MAP Check 3 Vacuum
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dansensor® MAP Check 3 Vacuum EN User Guide 55
P/N 320598-G
11/2019
Alarm settings Setting of the sample gas measurement warning and alarm values.
You have the following options for each alarm:
Off Disables all alarms
On Enables alarm for the selected input
-O
2
High alarm and Low alarm (%)
-O
2
High warning and Low warning (%)
-CO
2
High alarm and Low alarm (%)
-CO
2
High warning and Low warning (%)
If alarm settings are as shown above, O
2
measurements between 0,3
and 0,7% are OK, measurements between 0,7 and 1,0% will result in
a high warning, an measurements above 1,0% will result in a high
alarm.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Buffer alarm settings Setting of buffer measurement warning and alarm values.
Buffer pressure is measured at every “EVAC” signal.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Extra sampling time Prolongs the sampling period. If the gassing time is very short the
gas can spread through the system during the prolonged sampling
period.
Recommended setting is 1 sec.
Product alarm settings
Ok
%O
2
1,0
0,7
0,3
0,1
28,5
27,0
23,0
21,5
%CO
2
Product alarm settings
Ok
%O
2
Off
Off
Off
Off
Off
Off
Off
Off
%CO
2
Buffer alarm settings
Buffer alarm settings
Ok
Ok
%O
2
%O
2
bar bar
Off
1,0
10,0
Off
0,7 8,0
Off
0,3 4,0
Off 0,1 2,0
Off Off 28,5
Off Off
27,0
Off Off
23,0
Off Off 21,5
%CO
2
%CO
2

Table of Contents

Related product manuals