Alarm settings Setting of the sample gas measurement warning and alarm values.
You have the following options for each alarm:
• Off Disables all alarms
• On Enables alarm for the selected input
-O
2
High alarm and Low alarm (%)
-O
2
High warning and Low warning (%)
-CO
2
High alarm and Low alarm (%)
-CO
2
High warning and Low warning (%)
If alarm settings are as shown above, O
2
measurements between 0,3
and 0,7% are OK, measurements between 0,7 and 1,0% will result in
a high warning, an measurements above 1,0% will result in a high
alarm.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Buffer alarm settings Setting of buffer measurement warning and alarm values.
Buffer pressure is measured at every “EVAC” signal.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Extra sampling time Prolongs the sampling period. If the gassing time is very short the
gas can spread through the system during the prolonged sampling
period.
Recommended setting is 1 sec.
Ok
1,0
0,7
0,3
0,1
%CO
2
Product alarm settings
Ok
%O
2
Off
Off
Off
Off
Off
Off
Off
Off
%CO
2
Buffer alarm settings
Ok
bar bar
1,0
10,0
Off
0,7 8,0
Off
0,3 4,0
Off 0,1 2,0
Off Off 28,5
Off Off
27,0
Off Off
23,0
Off Off 21,5
%CO
%CO