15.2. TECHNICAL MAINTENANCE
These instructions describe the maintenance procedures for the RX DC X-ray unit. These instructions apply to all the revisions of said equipment.
In order to ensure the operational safety and functional reliability of the equipment installed, at least once a year an authorized service technician must perform a full inspection of the device.
When taking measurements that require a multimeter, always use a calibrated digital multimeter.
All the following tests will be carried out. Customer should be notified prior to replacing parts.
Answer questions with yes (√) or no (−)
Check that all labels located
- on the wall-mounted cover,
- on the X-ray tube
- inside the collimator/collimators,
are intact, well attached and legible.
User Manual, section Identification nameplates
Check there are no external damages to the equipment which could that may
reduce protection against radiation.
User Manual, section Description of the RX DC x-ray unit
Pull out the collimator and panel stop ring, take off the screw cover caps and
loosen the screws that secure the lower cover. Check there is no oil leakage on
the tube-head.
Technical Manual, section X-ray Head
Check the electromechanical brake that locks/unlocks the movement of the
generator and adjust it if necessary – Hypersphere model units only –
Technical Manual, section Actuator unit
Power off the unit and remove the wall mounting cover. Disconnect the unit
from the main power supply and check the condition of the main power supply
cable. Replace it in case of damage. Connect it back making sure the safety
ground is securely connected. Install the wall mounting cover back again.
Technical Manual, section Wall mounted plate wiring connections
Check the battery level of the wireless remote control
User Manual, section Batteries and charge level indication
Check the remote control functionality: buttons must respond to interaction
User Manual, section Control Panel
Check the power switch verifying that the switch is working properly and the
main switch green light illuminates when the switch is in the on position.
Section Turning the x-ray unit on and off
Check proper functioning of X-ray generator indicator light - Hypersphere model
units only
User Manual, section X-ray generator indicator light
Check the exposure buzzer during a trial x-rays emission
User Manual, section Control Panel
Verify that exposure is immediately interrupted when X-rays button is released
Check the scissors arm balance and adjust it if necessary
Technical Manual, section Balancing the double pantograph arm
Perform a calibration of the X-ray tube housing
Technical Manual, section Calibrating the X-ray Head
The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed over the unit in fully functional condition.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!.