RADIOGENIC TUBE F.S.=0.4 mm
- Radiogenic tube: TOSHIBA D-041.
- Focal spot: 0.4 mm in compliance with IEC 336 /
1993.
- Tolerance for position of the focal spot along the
reference axis: ± 2%.
- Nominal high voltage and maximum allowable current:
(65KV, 7mA) ± 10%.
- Anode construction material: Tungsten (W).
- Anode inclination: 12.5°
- Anode thermal load: 4.3 KJ (6 KHU).
- Maximum continuous heat dissipation: 100 W.
- Operating cycle: 1:60 (1 second exposure - 60
seconds pause time).
RADIOGENIC TUBE F.S.=0.7mm
- Radiogenic tube: TOSHIBA D-0711.
- Focal spot: 0.7mm in compliance with IEC 336 / 1993.
- Tolerance for position of the focal spot along the
reference axis: ± 2%.
- Nominal high voltage and maximum allowable current:
(65KV, 7mA) ± 10%.
- Anode construction material: Tungsten (W).
- Anode inclination: 16.0°.
- Anode thermal load: 7.0 KJ (10 KHU).
- Maximum continuous heat dissipation: 210 W.
- Operating cycle: 1:60 (1 second exposure - 60
seconds pause time).