April 27, 2015 10-11
Electromagnetic Compatibility (EMC)
(a*) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the Nicolet EEGwireless32/64 amplifier is used exceeds the
applicable RF compliance level above, the Nicolet EEGwireless32/64 amplifier should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the Nicolet EEGwireless32/64 amplifier.
(b*) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.