April 27, 2015 5-7
Using the Nicolet system to setup and acquire EEG
Check the
impedance
NOTE: Skip step 13 if the Startup in Impedance mode checkbox was checked
earlier (Tools > Options > Acquisition tab). Checking this box causes the
Impedance window to appear automatically when you open the Recorder software.
Please refer to the Nicolet Brand System Software Reference Guide, Miscellaneous
Quick Steps
chapter, Acquisition tab for additional information.
NOTE: If you disconnect the Headbox from the Amplifier during an Impedance test
and then reconnect it, the LEDs on the Headbox will not reappear. You must restart
the Impedance test to illuminate the LEDs again.
3. Select Impedance to display the Impedance Test window.
- or -
Select Acquisition > Impedance.
4. Select the acceptable impedance range:
Click on the Threshold show menu button and then select the desired
threshold value.
- or -
Enter a custom value.